International Standard
ϲʿ 19211:2024
Anaesthetic and respiratory equipment — Fire-activated oxygen shut-off devices for use during oxygen therapy
Reference number
ϲʿ 19211:2024
Edition 1
2024-08
Preview
ϲʿ 19211:2024
84738
No disponible en 貹ñDZ
Publicado (徱ó 1, 2024)

ϲʿ 19211:2024

ϲʿ 19211:2024
84738
Idioma
Formato
CHF 96
Convertir Franco suizo (CHF) a

Resumen

This document specifies requirements for fire activated oxygen shut-off devices that stop the flow of oxygen in respiratory therapy tubing when activated by fire.

NOTE 1 Typical arrangements for fire activated oxygen shut-off devices are shown in Annex C.

NOTE 2 Respiratory therapy tubing is covered by ϲʿ 17256.

NOTE 3 Use of fire activated oxygen shut-off devices in medical devices or accessories is not mandated in this document.

The fire activated oxygen shut-off devices specified in this document are not suitable for use with oxygen therapy systems with flows in excess of 20 l/min).

NOTE 4 There is rationale for this clause in A.2.

The requirements in this device-specific standard take precedence over any conflicting requirements in the general standard for airway devices (ϲʿ 18190). All the common requirements that appear in the general standard for airway devices have been removed from this document.