Abstract
This document specifies a method for calibrating the kinetic energy scales of Auger electron spectrometers with an uncertainty of 3Â eV, for general analytical use in identifying elements at surfaces. This document also specifies a method for establishing a calibration schedule.
It is applicable to instruments used in either direct or differential mode, where the resolution is less than or equal to 0,5Â % and the modulation amplitude for the differential mode, if used, is 2Â eV peak-to-peak. It is applicable to those spectrometers equipped with an inert gas ion gun or other method for sample cleaning and with an electron gun capable of operating at 4Â keV or higher beam energy.
General information
-
Status: PublishedPublication date: 2024-07Stage: International Standard published [60.60]
-
Edition: 3Number of pages: 11
-
Technical Committee :Ïã¸ÛÁùºÏ²Ê¿ª½±/TC 201/SC 7ICS :71.040.40