Abstract
This document specifies the properties and performance of sterile single-use hypodermic syringes with an auto-disable syringe feature intended to deliver a fixed dose of vaccine immediately after filling. The syringes can be made of plastic, rubber or other materials and can be with or without needle and needle protection feature.
This document does not specify the design of the auto-disable syringe feature.
This document is not applicable to syringes for use with insulin (covered by Ïã¸ÛÁùºÏ²Ê¿ª½± 8537), syringes for use with power-driven syringe pumps (covered by Ïã¸ÛÁùºÏ²Ê¿ª½± 7886-2), reuse prevention syringes (covered by Ïã¸ÛÁùºÏ²Ê¿ª½± 7886-4) or syringes designed to be prefilled (covered by the Ïã¸ÛÁùºÏ²Ê¿ª½± 11040 series). It does not address compatibility with injection fluids/vaccines.
General information
-
Status: PublishedPublication date: 2020-05Stage: International Standard published [60.60]
-
Edition: 2Number of pages: 15
-
Technical Committee :Ïã¸ÛÁùºÏ²Ê¿ª½±/TC 84ICS :11.040.25